close
close

Rosemount

Micro Motion H-Series Hygienic Coriolis Meter Consisting of H100S Sensor and 1500 Transmitter

Write a Review
SKU:
micro-motion-pkg-ft-h100s-1500
Request a Quote
Mass flow accuracy (Liquid): ±0.15% of rate (standard)±0.10% of rate (optional) Mass flow repeatability (Liquid): ±0.05% of rate Volume flow accuracy (Liquid): ±0.25% of rate (standard)±0.15% of rate (optional) Volume flow repeatability (Liquid): ±0.05% of rate Mass flow accuracy (Gas): ±0.50% of rate Mass flow repeatability (Gas): ±0.25% of rate Density accuracy (Liquid): ±0.002 g/cm3 (±2.0 kg/m3) (standard)±0.001 g/cm3 (±1.0 kg/m3) (optional)±0.0005 g/cm3 (±0.5 kg/m3) (optional) Temperature accuracy: ±1 °C ±0.5% of reading Sensor maximum working pressure: 1,450 psig (100 barg) Meter approvals, calibrations, and certifications: •CSA•IECEx•ATEX Available Outputs: Model 1500 : for applications requiring one flow rate variable ( in milliamp and a frequency/pulse outputModel 2500 : for applications requiring simultaneous monitoring of multiple flow variables in combinations of outputs (mill Housing: Polyamide PA 6.6 Power: • Minimum 19.2 to28.8 VDC• 6.3 watts• Meets Installation (Overvoltage) Category II, Pollution Degree 2 requirements Input/output signals: • One active 4–20 mA output• One active frequency output HART: Bell 202 signal is superimposed on the primary milliamp output Mounting options: Remote-mount control room only• Mounted on 35 mm rail• Rail must be grounded Application types: • Continuous process control• Two-phase flow/entrained gas• Filling &dosing• Meter verification Hazardous approvals: Suitable for non-hazardous environments only
Worldwide Shipping
Tailored Quotation
Secure Shopping
24/7 Customer Support
Icon Hot
Icon Hot