Mass flow accuracy (Liquid): ±0.10% of rate (standard)±0.05% of rate (optional)
Mass flow repeatability (Liquid): ±0.05% of rate (standard)±0.025% of rate (optional)
Volume flow accuracy (Liquid): ±0.10% of rate (standard)±0.05% of rate (optional)
Volume flow repeatability (Liquid): ±0.05% of rate (standard)±0.025% of rate (optional)
Mass flow accuracy (Gas): ±0.35% of rate
Mass flow repeatability (Gas): ±0.20% of rate
Density accuracy (Liquid): ±0.0005 g/cm3 (±0.5 kg/m3) (standard)±0.0002 g/cm3 (±0.2 kg/m3) (optional)
Density repeatability (Liquid): ±0.0002 g/cm3 (±0.2 kg/m3) (standard)±0.0001 g/cm3 (±0.1 kg/m3) (optional)
Temperature accuracy: ±1 °C ±0.5% of reading
Sensor maximum working pressure: 2,855 psig (197 barg)
Meter approvals, calibrations, and certifications: •UL•CSA•IECEx•NEPSI•ATEX•Ingress Protection•EMC effects
Output variables: Mass & volume flowNet product content / flowTemperatureDensityConcentration
Local display: Multi-line LCD panelBacklit LCD, 128 × 128 pixel, bit mapAdjustable contrastNon-glare, chemical-resistant acrylic lensLarge push buttons with tactile feedbackSoftware-defined function keys
Power: AC option:85 to 265 VAC50/60 Hz, 30 VAComplies with European low-voltage directive 2006/95/EC per IEC 61010-1, and Installation (Overvoltage) Category II, Pollution Degree 2DC option:18 to 30 VDC18 wattstypical, 25 watts maximum
Input/output signals: Two active 4–20 mAThree discrete outputsOne 2-wire frequency outputOne 2-wire frequency inputTwo discrete momentary inputs
HART/WirelessHART: Bell 202 signal is superimposed on the primary milliamp output
Mounting options: Remote field-mount
Application types: Batch controllerCustody transferTwo-phase flow/entrained gasMeter verification
Hazardous approvals: ULCSAATEXNEPSI