Pipe Clamp Materials: ASTM 304 SST (1.4301), Duplex F51 (1.4462), Carbon steel (1.0037)
Lead Wires: PTFE insulated, silver-coated copper wire
Insulation Resistance: 1,000 MΩ minimum insulation resistance when measured at 500 V dc at room temperature
Calibration Options: Class A Sensor from -50 to 300 °C (-58 to 572 °F)
Certifications/Approvals: Hazardous location, see full specs for complete list of certifications
Input: Dual and single sensor capability with universal sensor inputs (RTD, T/C, mV, ohms)
Output Signal: 4-20 mA /HART™ protocol, FOUNDATION™ fieldbus protocol, PROFIBUS™ PA protocol
Housing: Field mount, DIN A head mount or rail mount
Display/Interface: LCD Display, LCD Display with Local Operator Interface (LOI), or LED
Diagnostics: Basic diagnostics, Hot Backup™ capability, sensor drift alert, thermocouple degradation, min/max tracking
Calibration Options: Transmitter-sensor matching (Callendar-Van Dusen constants), 5-point calibration
Certifications/Approvals: SIL 2/3 certified to IEC 61508 by an independent 3rd party, hazardous location, marine type, see full specs for complete list of certifications