Mass flow accuracy (Liquid): ±0.15% of rate
Mass flow repeatability (Liquid): ±0.05% of rate
Volume flow accuracy (Liquid): ±0.25% of rate
Volume flow repeatability (Liquid): ±0.05% of rate
Mass flow accuracy (Gas): ±0.50% of rate
Mass flow repeatability (Gas): ±0.05% of rate
Density accuracy (Liquid): ±0.002 g/cm3 (±2.0 kg/m3)
Density repeatability (Liquid): ±0.0005 g/cm3 (±0.5 kg/m3)
Temperature accuracy: ±1 °C ±0.5% of reading
Sensor maximum working pressure: 1,450 psig (100 barg)
Meter approvals, calibrations, and certifications: •UL•CSA•ATEX•ASME•3-A•USDA•EHEDG
Output variables: •Mass & volume flow •Net product content / flow •Temperature •Density •Concentration
Local display: None
Power: •Minimum 19.2 to 28.8 VDC •6.3 watts •Meets Installation (Overvoltage) Category II, Pollution Degree 2 requirements
Input/output signals: •One active 4–20 mA output •One active frequencty output
HART: Bell 202 signal is superimposed on the primary milliamp output
Mounting options: •Mounted on 35 mm rail •Rail must be grounded
Application types: •Continuous process control •Two-phase flow/entrained gas •Meter verification
Hazardous approvals: Suitable for non-hazardous environments only